Testo Smart Connect, revisited
If you have just read the CCleaner post, this one is the other kind of story.
A few days after I published a post gently criticising the Testo Smart Connect migration, the team responsible for it got in touch. Not to object. To say thank you.
This is not how these things usually go, so it is worth writing down.
The email
The message came from Pascal Hauer, the Product Manager responsible for the Saveris 2 to Smart Connect migration. A colleague had shared my post with him, and he wrote personally – to thank me for what he described as candid, constructive feedback, to confirm that the two commissioning issues I had described were real and that fixes were actively being rolled out, and to say that posts like mine help the team prioritise the rough edges that matter most to actual users.

Older generation Testo 162 T2 temperature logger, showing clearly the new display layout after the update to Testo Smart Connect.
© RudolphLAB, 2026
He had also, notably, read the whole thing. He picked out the story of the rescued strain collection – the reason I persist with continuous temperature monitoring in the first place – and noted that it was exactly the kind of outcome the system exists to protect.
And then he asked whether I would be willing to spend fifteen or twenty minutes on a call, to talk about where the logger onboarding experience could be improved, and to address anything still outstanding on my end.
I said yes.
The call
It was, I have to report, an entirely delightful conversation.
Some of it was context I had not known. As it turns out SLS had recommended Testo to me relatively early in their own history of selling these loggers, which lent a certain shared-history quality to the discussion. But then the system had been quietly working in the background for years, so I had spent little effort learning more about the company. It was interesting and quite revealing and I had no idea about the numbers of customers they have, which makes this interaction even more remarkable. However, most of the call was Pascal being genuinely interested in how a real user had experienced the migration, rather than how the team imagined a user would.
I need to make a point of being perfectly honest about the limits of my own experience: I have recently upgraded the entire infrastructure around the loggers, which meant I spent considerable time looking into the new SIM router interface, establishing connections with the old loggers and overall getting the system up and running again – which is precisely why, when it came to connecting the two new loggers, I never consulted the documentation. Thinking they would behave like my older loggers was precisely why I fell into the commissioning traps I wrote about. A genuinely new user might read the manual and avoid them entirely. But then again, a genuinely new user might not, and the value of a design is in how it behaves for the person who does not read the instructions.
Which led to a clean and simple solution for the first problem: The new loggers ship with a protective foil over the display. The commissioning trap – that the device will not connect until a temperature probe is attached – could be headed off entirely by printing a single line on that foil: please attach the temperature probe before connecting the logger to the network. The foil has to be peeled off anyway. The warning arrives at exactly the right moment and then removes itself. No cost, no manual, no support call. Simple and efficient.
The alert that is not there yet
The second issue is more serious, and I have already had a separate exchange with James in UK technical support about it, whose help throughout the entire experience has been truly excellent.
On the old Saveris 2 platform, I received a notification whenever a logger that had lost communication came back online – a confirmation that normal operation had resumed. Smart Connect, as currently deployed, does not send this. The industry term, which Pascal supplied and I promptly forgot until reminded, is a return-to-normal notification.
This sounds like a minor convenience. It is not, and the reason is worth spelling out, because it is a genuine gap in a safety-critical system.
Some time ago we had a Wi-Fi interruption on a Friday evening. The logger correctly reported the loss of communication. I saw the alert – and then, over the weekend, forgot about it. The freezer itself then failed just a few hours later, while the connection-interruption alert was still active. And here is the problem: because the logger disconnected from the Wi-Fi network (and had reported this correctly), no new temperature alarm could be submitted when the freezer actually failed. This lack of connection, correctly flagged, made a second and far more important one impossible. By the time we realised, the contents had thawed.
Before this becomes too neat a moral: the temperature alert email likely would not have helped. The freezer stopped working on a Friday evening at a time when I very likely would have stopped monitoring emails. So, even if the alert email would have come through correctly, I very likely would have missed it. I would have found it the following morning, about 12 h after the freezer failed, and approximately 11 h after the temperature alert would have been issued. As this was a small underbench freezer, by this time the temperature had already risen to +1 °C. The timing was simply working against us.
Still, since then I have watched connection-interruption alerts closely, and made a point of confirming that communication has been restored. On the old platform that was effortless, because the confirmation arrived by email. Without it, I have to remember to log into the dashboard and check – which is exactly the kind of manual vigilance that fails on a Friday night.
I raised this with James, asking whether I had simply missed a setting. His reply was refreshingly direct: no, I had not missed anything; the feature is not yet on the new platform, but it has been identified as a key feature and will be incorporated in updates rolling out later during 2026, with no exact date yet. Pascal, on the call, went further and told me it sits second in the development pipeline.
No deflection. No "many variables affect the working life of a notification." Just: yes, it is missing, yes, it matters, we know, and here is roughly when it will be fixed.
We also talked about the fact that what used to be entirely free, rather astoundingly, is now a subscription service with a fee attached. Coming from the free version, and having to pay for this out of my own pocket, it is a slightly unfortunate development, but I have enjoyed the entirely free service for many years, and given the value of what it protects it is a genuinely low price to pay. I said it before, and I say it again: how they managed to keep the service free for so many years in the first place is rather surprising, as the infrastructure costs the company money, of course.
Plus, there is one thing that I now get as a paying customer: I get combined temperature profiles. In the free version I had to select every temperature logger separately. Genuinely a small thing to do. But, having now all three temperature profiles in one window, visible at a quick glance, is definitely worth more than I anticipated.
Why this matters more than a freezer
I cannot overstate how much this differs from other experiences I have documented.
Testo cannot, of course, personally contact every customer who writes about them. My post evidently struck a note, they judged it fair, and they reached out. For me, as an individual customer, that is a remarkable thing. But what elevates it beyond a nice gesture is what happened next: they did not sugar-coat the missing features, they did not defend the rough edges, and they did not treat my write-up as a complaint to be managed. They treated it as a conversation to be had, and arrived with solutions. Even if my points had been unjustified, this approach helps. I have learned this through more student conversations than I care to count. Sometimes students perceive work they have done differently to the grade they have been awarded. My personal experience is that a conversation, done in earnest and with proper listening to the student's points, always made the complaint go away, even if the result never changed. Taking the complaint seriously and then having a conversation about why exactly something was rated a certain way helped understanding and left all parties happier at the end, even if nothing changed. And I admit freely: I have complained about things which, from the manufacturer's side, would have looked probably silly. A conversation still would have helped.
It is instructive to set this against three other stories on this blog. When I bought a FileMaker Pro upgrade that was obsolete within weeks, Claris saw nothing wrong with the situation and defended it as correct procedure. When three Liebherr freezers failed in a row within three years, the technical team told me that many variables affect the working life of a unit and offered me an out-of-warranty service visit. In both cases the underlying message was: this is your problem, not ours. The most recent is CCleaner, whose support desk answered a fault report over four months with replies that took no account of what I had already told them. There the message was something stranger: nobody was reading.
Testo’s message was the opposite of both. As a customer, my concerns and experiences felt taken seriously.
I might confirm on this blog whether that return-to-normal alert arrives later in 2026 as promised. But I am not especially worried. When a company tells you plainly what is broken and where it sits in the queue, you tend to believe them.